Glossary
of Terms in Mass Spectrometry
Acquisition Time (Mass Spectrometer Acquire Time)
The
total mass spectrometer analysis time. During
this time data are stored on the hard disk drive.
Array
Detectors
Solid
state devices which detect ions impacting over a region.
Atmospheric Pressure Chemical Ionization (APCI)
A
very soft chemical ionization technique [but not as soft
as electrospray ionization (ESI)] carried out in an ion
source operating at atmospheric pressure. Electrons from
a corona discharge initiate the process by ionizing the
mobile phase vapor molecules. A reagent gas forms, which
efficiently produces positive and negative ions of the analyte
through a complex series of chemical reactions.
APCI
is typically used to analyze small molecules with molecular
weights up to about 2000 u. APCI is a very robust ionization
technique. It is not affected by minor changes in most variables,
such as changes in buffer or buffer strength.
Auxiliary Gas
The
outer-coaxial gas (nitrogen) that assists the sheath (inner-coaxial)
gas in dispersing and/or evaporating sample solution as
the sample solution exits the APCI or ESI nozzle. The auxiliary
gas also sweeps solvents out of the API source to help keep
the interior of the ion source as dry as possible. The auxiliary
gas is used especially at higher solvent flow rates.In the
static NSI source, the auxiliary gas can be used to provide
backpressure. The backpressure required to initiate the
spray is often higher than that required to maintain the
spray. (Auxiliary gas is not used with the dynamic NSI source.)
Calibration
Mass
calibration: A procedure performed to ensure that the mass spectrometer mass assignments
are accurate. You perform a mass calibration with a compound
that has a well-known mass spectrum.
Quantitative calibration: The process of determining how the LC/MS system responds
to different compounds. Quantitative calibration involves
designing a processing method and performing calibration
experiments with calibration standards containing known
quantities of analyte.
Calibration, Mass Spectrometer
A
procedure performed to ensure that the mass spectrometer
mass assignments are accurate. You perform a mass calibration
with a compound that has a well-known mass spectrum.
Calibrating the mass spectrometer involves setting the electron
multiplier gain, calibrating mass and resolution for various
scan modes, and calibrating RF voltage parameters to achieve
optimum mass spectral performance.
Centroid Data
Data
used to represent mass spectral peaks in terms of two parameters,
the centroid (the weighted center of mass) and the intensity.
The data is displayed as a bar graph. Centroid data differ
from profile data. In profile data format, the mass spectral
peak is represented as a series of data points, each with
its own intensity value. In centroid data format, the transformed
representation of this mass spectral peak consists of the
centroid of this series of data points and a single intensity
value that represents the sum of the intensities of the
data points.
Centroiding
A
method used to improve mass spectral data quality, get better
mass assignments, and reduce data file size. Profile data,
in which many points are used to delineate a mass spectral
peak, is converted into mass-centroided data by a data compression
algorithm. The centroided mass peak is located at the weighted
center of mass of the profile peak. The normalized area
of the peak provides the mass intensity data.
Charge State
The
imbalance between the number of protons (in the nuclei of
the atoms) and the number of electrons that a molecular
species (or adduct ion) possesses. If the species possesses
more protons than electrons, its charge state is positive.
If it possesses more electrons than protons, its charge
state is negative.
For
positive ion ESI, medium to large ions frequently have charge
states of greater than +1 by the addition of protons. For
negative ion ESI, medium to large ions frequently have charge
states of greater than -1 by the loss of protons. For positive
ion APCI, ions typically have a charge state of +1 by the
addition of a proton. For negative ion APCI, ions typically
have a charge state of -1 by the loss of a proton or electron
capture.
Chemical Ionization (CI)
The
formation of new ionized species when gaseous molecules
interact with ions. The process can involve transfer of
an electron, proton, or other charged species between the
reactants. When a positive ion results from chemical ionization,
the term can be used without qualification. When a negative
ion results, the term negative ion chemical ionization can
be substituted.
Chemical Ionization (CI) Plasma
The
collection of ions, electrons, and neutral species formed
in the ion source during chemical ionization. Chemical ionization
is a two-step process. First, CI reagent ions are formed
in the CI plasma. Then, sample ions are formed by gas phase
reactions with the CI reagent ions.
Chromatogram
The
graphical representation of a chemical separation, such
as liquid chromatography, obtained from an analytical instrument
called a chromatograph. The result of plotting detector
response versus time.
Chromatography Filter
The
chromatography filter is a sophisticated noise reduction
algorithm that smoothes peak profiles at low sample concentrations
in chromatographic data. These peak profiles integrate more
consistently than those acquired under normal data acquisition
conditions.
Collision Energy
The
translational kinetic energy (TKE) of a (moving) projectile
ion when it interacts with a (stationary) target neutral
species, such as a collision gas molecule. Collision
energy is an important factor in the collision induced dissociation
(CID) process.
Collision-Induced
Dissociation (CID)
An
ion/neutral process in which a (fast) projectile ion is
dissociated as a result of interaction with neutral target
species. Dissociation is brought about by the conversion
of part of the translational kinetic energy of the ion into
internal energy in the ion during the collision. [Also
referred to as collision-activated dissociation (CAD)].
Data
Systems
Computer
based instruments which process and record the detector
signals as well as providing control for all the processes
involved in operating the instrument.
Divert/Inject Valve
A
valve on the mass spectrometer that can be plumbed as a
divert valve or as a loop injector. As a loop injector,
you can load the valve and then inject the contents of the
valve into the ion source of the mass spectrometer. As a
divert valve, the valve directs the effluent from an LC
into the ion source of the mass spectrometer for the analysis
of samples, or it diverts the LC effluent to a waste container.
EI
(Electron Ionisation)
Ionisation
in the gas phase by interaction with a beam of electrons.
Electron
Multipliers or Photomultipliers
Detection
of ions by production of an amplified electric current,
either as a consequence of direct impact of the ion on a
surface or following production of a photon from a phosphor
followed by a photomultiplier tube.
Electrospray Ionization (ESI)
A
type of atmospheric pressure ionization that is currently
the softest ionization technique available to transform
ions in solution into ions in the gas phase. The analyte
solution is forced through a narrow-bore spray capillary
(needle) to which a potential has been applied. The high
potential causes the emerging solution to disperse into
a fine spray of charged droplets. As the solvent evaporates
away, the droplet size shrinks and the charge concentration
on the surface increases. Eventually, the coulombic repulsion
overcomes the droplet's surface tension. The droplet explodes
and forms a series of smaller, lower-charged droplets. This
process continues until the analyte ion escapes the droplet
(ion desorption) or all of the solvent evaporates to leave
the bare ion (ion evaporation). This ionization technique
produces ions in multiple charge states. As a result, a
mass spectrum can have multiple peaks for ions of the same
mass.
Many
samples that previously were unsuited for mass analysis
(for example, heat-labile compounds or high molecular weight
compounds) can be analyzed by the use of ESI. ESI can be
used to analyze any polar compound that makes a preformed
ion in solution, including adduct ions.
ESI Spray Current
The
flow of charged particles in the ESI source. The voltage
on the ESI spray needle supplies the potential required
to ionize the particles. The ESI spray current varies; the
ESI spray voltage is set. A type of electrospray ionization
(ESI) that accommodates very low flow rates of sample and
solvent on the order of 1 to 20 nL/min (for static
nanospray) or 100 to 1000 nL/min (for dynamic nanospray).
Nanospray allows for ESI analysis of very small sample volumes
(approximately 1 µL). This in turn allows the use of
more concentrated solutions from small amounts of sample
and/or longer analysis times because of the very low flow
rates.
Static
nanospray permits continuous analysis of small analyte solution
volumes over an extended period of time. It is particularly
useful for obtaining detailed structural information using
MS/MS for purified samples or those present in a simple
mixture. Sample is loaded into a metal-coated glass capillary
with a tip internal diameter that ranges from 1 to 4 mm.
A voltage is applied to the metal coating on the tip and
an air-filled syringe provides a constant backpressure to
initiate and maintain electrospray. After electrospray has
begun, the electric field and the solvent properties allow
the flow to be self-regulated.
Dynamic
nanospray employs microscale capillary columns to separate
the analytes in complex mixtures. The sample is loaded onto
a column using an injection valve or a gas pressure vessel.
The mixture components are then eluted by a solvent gradient
and pumped through the emitter.
FT
(Fourier Transform)
The
mathematical operation used to convert the signals detected
in an ICR mass spectrometer to a set of m/z values.
Full
Scan Type
A
scan type that provides a full mass spectrum of each analyte
or parent ion. With the full scan type, the mass analyzer
is scanned from the first mass to the last mass without
interruption in a given scan time. The full scan type
can be employed in the following three cases:
- Single-stage
(MS) full scan type.
- Two-stage
(MS/MS) full scan type.
- Multi-stage
(MS^n) full scan type.
GC/MS
(Gas Chromatography/Mass Spectrometry)
The
linking of mass spectrometry to a gas chromatographic system.
The gaseous effluent from a capillary GC column can be introduced
directly into the mass spectrometer source.
Ionization Mode |
Scan Filter Abbreviation |
|
atmospheric pressure chemical ionization |
APCI |
|
chemical ionization |
CI |
|
electron ionization |
EI |
|
electrospray |
ESI |
|
fast
atom bombardment |
FAB |
|
field
desorption |
FD |
|
glow
discharge |
GD |
|
matrix-assisted laser desorption/ionization |
MALDI |
|
nanospray |
NSI |
|
thermospray |
TSP |
ICP-MS
(Inductively Coupled Plasma Mass Spectrometry)
Ionization
of elements in a high temperature plasma.
Ion Polarity Mode
The
mass spectrometer can operate in either of two ion polarity
modes: positive ion, in which positive ions are analyzed,
and negative ion, in which negative ions are analyzed.
Ion Source
The
source of ions (molecular ions and fragment ions, or in
an MS/MS experiment, parent ions). Your mass spectrometer
can employ any of several ion sources: an atmospheric pressure
chemical ionization (APCI) source, an electrospray ionization
(ESI) source, or a nanospray ionization (NSI) source.
Ion-Trap
Ions
are pulsed into a three dimensional device where they are
stored by a quadrupole field. They can then be ejected at
defined m/z values by the application appropriate secondary
electric fields.
Isotopes
Most
elements are composed of a mixture of isotopes. These will
be separated in a mass spectrometer. Atoms or molecules
containing such elements will display a cluster of ions
reflecting the isotopic composition.
LC/MS
(Liquid Chromatography/Mass Spectrometry)
The
linking of the effluent from a liquid chromatographic system
to a mass spectrometer.
MSn
A
term used in the case of MS/MS experiments with ion traps
where product ion analysis is extended over a number of
generations.
MALDI
(Matrix Assisted Laser Ionisation)
Ionization
by effect of illumination with a beam of laser generated
light onto a matrix containing a small proportion of analyte.
Mass
Analyzer
A
devise that separates ionic species according to the mass-to-charge
ratios of the ions.
Mass Range Chromatogram
A
type of chromatogram in which all ions from one or more
specified mass spectrometer mass ranges are summed and plotted
as a function of time.
A
plot of the ion intensity of a specified mass or masses
[Y-axis] versus the corresponding scan number or time [X-axis].
The ion intensity is the sum of all ion intensities for
data in a selected mass range or mass ranges of each scan.
Mass Spectrometer
An
instrument that ionizes sample molecules and then separates
the ions according to their mass-to-charge ratio (m/z).
The resulting mass spectrum is a characteristic pattern
for the identification of a molecule.
All
mass spectrometers have four common components:
- A system for introducing the sample to be analyzed into
the instrument
- A system for ionizing the sample
- An accelerator that directs the ions into the measuring
apparatus
- A system for separating the constituent ions and recording
the mass spectrum (a plot of ion abundance versus ion
mass-to-charge ratio) of the sample.
Abbreviated: MS.
Mass
Spectrometry
This is the technique in which an instrument is employed to produce
ions from atoms or molecules (the source) which are
then separated according to their charge-to-mass-ratios
(the analyzer) and detected.
Mass Spectrum
A
spectrum obtained when ions are separated according to their
mass-to-charge ratios (m/z). A graphical representation
(plot) of measured ion abundance versus mass-to-charge ratio.
The mass spectrum is a characteristic pattern for the identification
of a molecule and is helpful in determining the chemical
composition of a sample.
m/z
Mass
to charge ratio. An abbreviation used to denote the quantity
formed by dividing the mass of an ion (in u) by the number
of charges carried by the ion. For example, for the ion
C7H72+, m/z = 45.5.
MS/MS
Mass
spectrometry / mass spectrometry or tandem mass spectrometry.
An analytical technique that involves two stages of mass
analysis. In the first stage, ions formed in the ion source
are analyzed by an initial analyzer. In the second stage,
the mass-selected ions are activated by collision with a
neutral gas, and the resultant ionic fragments are mass
analyzed by a second analyzer.
MS
Scan Mode
The
scan power equal to 1, where the scan power is the power
“n” in the expression MS^n.
The
MS scan mode corresponding to the one stage of mass analysis
in a single-stage full scan experiment or a selected ion
monitoring (SIM) experiment.
MS/MS
Scan Mode
The
scan power equal to two (2), where the scan power is the
power “n” in the expression MS^n. The MS/MS scan
mode corresponding to the two stages of mass analysis in
a two-stage full scan experiment or a selected reaction
monitoring (SRM) experiment.
Negative Ion
An
atom, radical, molecule, or part of a molecule that has
one or more electrons than it has protons, and thus has
a negative charge. The charge state (-1, -2, -3, etc.) depends
upon the number of electrons in excess of the number of
protons.
NICI
(Negative Ion Chemical Ionization)
Ionization
as a consequence of capture of an electron in the gaseous
phase.
Neutral
Loss Map
The
neutral loss mass is [parent mass – product mass].
Octapole
An
octagonal array of cylindrical rods that acts as an ion
transmission device. An RF voltage and dc offset voltage
applied to the rods create an electrostatic field that transmits
the ions along the axis of the octapole rods. (Also called
an octapole rod assembly.)
Parent Ion
An
electrically charged molecular species that can dissociate
to form fragments. The fragments can be electrically-charged
or neutral species. A parent ion can be a molecular ion
or an electrically charged fragment of a molecular ion.
Parent ions are produced in the API source. Also called
a precursor ion.
Parent Mass
The
mass-to-charge ratio of a parent ion. The location of the
center of a target parent-ion peak in mass-to-charge ratio
(m/z) units.
Peak Width
The
distance across a peak measured at a selected peak-height
level.
For
chromatograms:
The time in minutes between peak start and peak end
using the current peak integration settings
For
mass spectra:
The number of mass units across a peak measured at a designated
peak-height level. The peak-height level is usually specified
as a percentage of the maximum peak height.
Positive Ion
An
atom, radical, molecule, or part of a molecule that has
one or more fewer electrons than it has protons, and thus
has an electrically positive charge. The charge state (+1,
+2, +3, etc.) depends upon the number of protons in excess
of the number of electrons.
Product Ion
An
electrically charged product of reaction of a selected parent
ion. In general, product ions have a direct relationship
to a particular parent ion and can correlate to a unique
state of the parent ion. The reaction need not necessarily
involve fragmentation. It could, for example, involve a
change in the number of charges carried. Thus, all fragment
ions are product ions, but not all product ions are necessarily
fragment ions. Product ions are formed in the collision
cell.
Formerly
called a daughter
ion.
Product Mass
The
mass-to-charge ratio of a product ion. The location of the
center of a target product-ion peak in mass-to-charge ratio
(m/z) units.
Formerly
called the daughter
ion mass.
Profile Data
Data
representing mass spectral peaks as point-to-point plots,
with each point having an associated intensity value. Profile
data are different from centroid data. In profile data format,
the mass spectral peak is represented as a series of data
points, each with its own intensity value. In centroid data
format, the transformed representation of this mass spectral
peak consists of the centroid of this series of data points
and a single intensity value that represents the normalized
sum of the intensities of the data points.
Profile
Scan Data Type
A
scan data type that shows the shape of the peaks in the
mass spectrum.
Quadrupole(Q)
The
application
of a combination of DC and AC voltages to four parallel
rods creates a filter through which, on application of the
appropriate voltages, only ions of any one defined m/z value
can be transmitted. If only an Rf field is applied all ions
are transmitted (q)
Relative
Collision Energy
A
measure of collision energy given as a percent of the maximum
value. The term relative collision energy is used
in for two cases: octapole region source CID and mass
analyzer CID.
For
octapole region ion source CID: The relative collision
energy varies from 0.1% to 100% for octapole region ion
source CID. (0 to 100% relative collision energy corresponds
to 0 to-100Vdc of octapole offset voltage for positive ions
and 0 to+100Vdc for negative ions.)
For
mass analyzer CID: The relative collision energy varies
from 0 to 100% for mass analyzer CID. (0 to 100% relative
collision energy corresponds to 0 to 5V peak-to-peak of
resonance excitation RF voltage.)
Resolution
A
term which defines the performance of an analyzer. This
parameter describes the degree to which two ions of adjacent
m/z can be distinguished from each other with less than
a defined amount of overlap. The latter is often quoted
as 10% for sector instruments and 50% for quadrupoles. For
isolated peaks the same term can be calculated from the
width of the peak at defined heights.
Scan Event
A
mass spectrometer scan that is defined by choosing the necessary
scan parameter settings. Multiple scan events can be defined
for each segment of time.
Scan Power
The
power n
in the expression MSn. The number of stages of
mass analysis, expressed as MSn where n
is the scan power.
For
example, a scan power of n = 1 corresponds to
an MS1 (or MS) scan with one stage of mass analysis.
A scan power of n = 2 corresponds to an MS2
(or MS/MS) scan with two stages of mass analysis. A scan
power of n = 3 corresponds to an MS3
scan with three stages of mass analysis, and so on.
Scan Time
The
amount of time required to accomplish one scan, from the lowest
mass to the highest mass of a specified scan range. The scan
can be continuous (full scan type) or it can be in segments
(SIM and SRM, for example).
Selected Ion Monitoring (SIM) Scan Type
A
scan type with one stage of mass analysis and in which one
or more ions of particular mass-to-charge ratios are monitored.
In
the SIM scan type, the ions formed in the ion source are
stored in the mass analyzer. Then, ions of interest
for one or more mass-to-charge ratios are isolated and are
scanned out of the mass analyzer to produce a SIM mass spectrum.
(No data are acquired for other ions as they leave the mass
analyzer.)
Because
only ions of selected mass-to-charge ratio are monitored,
the selected ion monitoring scan mode generally provides
higher sensitivity than does the full scan mode.
Sheath Gas
The
inner coaxial gas (nitrogen), which is used in the API source,
that helps to nebulize the sample solution into a fine mist
as the sample solution exits the ESI or APCI nozzle. (Sheath
gas is not used with the NSI source.)
Source CID
Ion
source collision-induced dissociation (CID) is a technique
for fragmenting ions in an atmospheric pressure ionization
(API) source. Either of two regions can be used for
the ion source CID process: the capillary-skimmer
region and the octapole region.
Note:
Ion source CID is useful for some experiments. However,
its usefulness is limited compared to mass analyzer CID,
which provides true MS/MS and MS^n capabilities. Mass
analyzer CID provides greater specificity and higher efficiency
than does ion source CID. Therefore, it is usually
the method of choice.
Capillary-skimmer
region ion source CID: Ions emerge from the heated
capillary and expand into the capillary-skimmer region of
the API source with considerable translational kinetic energy.
This kinetic energy is further increased by applying greater
than usual voltages to the heated capillary and/or tube
lens, which increases the difference in potential between
these components and the skimmer (which is at ground potential).
The voltages are increased sufficiently to impart enough
kinetic energy to the ions so that, when they collide with
solvent or air molecules (collision gas) present in the
API source, the ions dissociate to form product ions.
Syringe Pump Flow Rate
The
rate of flow of liquid from the syringe pump (in microliters
per minute). The syringe pump flow rate can be set from
the Syringe Pump page of Instrument Setup. Typical syringe
pump flow rates are between 10 and 20 µL/min.
Tandem
Mass Spectrometry
The
use of two analyzers separated by a region in which ions
can be induced to fragment by transfer of energy (frequently
by collision with other molecules). (see MS/MS)
TOF
(Time-of-Flight)
The
ions formed are all accelerated by a pulsed potential down
an evacuated tube (drift region) and their time of arrival
at a detector is determined. This is a function of their
m/z values.
Tune,
Mass Spectrometer
Tuning
the mass spectrometer involves optimizing voltages, currents,
flows, and the like for the ion source parameters to achieve
the maximum mass spectral sensitivity and proper resolution.
The way the mass spectrometer is tuned affects all data
acquisition. The tuning of the various parameters is often
interdependent, with each parameter affecting the optimum
value of other parameters.