| Optical
System: Czerny-Turner(C-T) Type Monochromator |
| Wavelength
Range: 190nm ~ 900nm |
| Spectral
Bandwidth: 0.1 nm, 0.2 nm, 0.5nm, 1.0 nm, and 2.0 nm.
(5 steps, with automatic changeover) |
| Wavelength
Accuracy: +/-0.3 nm (automatic correction) |
| Wavelength
Reproducibility: 0.2 nm |
| Resolution:
over 0.3 nm |
| Baseline
Stability: 0.005A/30 min |
| Flame
Analysis |
| Sensitivity:
Cu 5µg/ml > 0.850 Abs |
| Burner
Head: Titanium Alloy Burner |
| Nebulizer:
High-efficiency Glass Nebulizer |
| Atomization
Chamber: Corrosion-resistant Material |
| Position
Adjustment: Automatic flame and furnace changeover,
Automatic flame burner optimum height setting |
| Safety:
Automatic ignition of mixing air-acetylene gas with
protection control |
| Graphite
Furnace Analysis |
| Characteristic
Mass (Cd): 0.8Pg |
| Temperature
Range: Ambient - 2650°C |
| Heating:
voltage feedback control when drying and ashing, optical
temperature control when atomizing |
| Heating
Program: up to 9 steps with choices of ramp, ladder
and full-power heating |
| Background
Correction |
| Deuterium
Lamp Background Correction: 1.0 Abs |
| Self-Reversal
Background Correction: 3.0 Abs |
| Data
Processing |
| Analytical
Method: flame AA, graphite furnace, hydride, and flame
emission |
| Determination
Method: calibration curves (1~3 order), standard addition
method |
| Repetition
Times: 1~20 times with calculations of average value,
SD and RSD |
| Result
Printout: output of parameters, data, spectra and calibration
curves |
| Mainframe |
| Light
Source: 8 lamps, 2 lamps simultaneously lit (with one
preheat) |
| Power
Supply: 220V/50Hz three-phase AC, 200W(mainframe), 5000W(graphite
furnace) |
| Mainframe
Size: 110cm x 50cm x 45cm, 75kg |
| Package:
Part 1: 1340x760x570mm 158Kg; Part 2: 540x410x750mm
42Kg; Part 3: 760x760x700mm 75Kg |
| Graphite
Furnace Power Supply System 50cm x 50cm x 45cm, 50kg |
| Features:
Fully Automatic Control |
- The
integration of flame burner and graphite furnace
in one compact unit enables automatic interchanging
between the flame mode and furnace mode using few
simple keystrokes.
- Eight-Lamp
turret for automatic positioning of each hollow
cathode lamp.
- Automatic
fuel gas(C2H2) flow rate adjustment to optimize
combustion efficiency in element analysis.
- Automatic
adjustment of burner height for optimized analysis
conditions.
- Automatic
Reflecting/Transmitting beamsplitter positioning
(to lead the light of deuterium lamp into optical
path) when the Deuterium Lamp Background Correction
function is selected.
- Automated
wavelength scanning and peak-seeking
- Automatic
switchover of spectral bandwidth (5 choices).
- Automatic
negative voltage and lamp current adjustment, automatic
AA and BG energy balance, automatic igniting and
interlock protection in case of flameout.
|
| Safety
Assurance |
- Flame
Monitoring: Constant monitoring. When accidental
flameout occurs, an alarm is given and the acetylene
flow is stopped immediately
- Pressure
Monitoring: Constant air pressure detection. When
abnormalities are detected, acetylene flow will
be automatically cut off for safety fire extinguishing.
The argon gas pressure monitor will disconnect the
electricity to the graphite tube when an absence
of the protective argon gas is detected.
- Cooling
water flow monitoring: The power supply to the furnace
will only be activated for heating when the cooling
water flow is determined enough to cool down the
furnace body.
- Graphite
Tube Fracture Protection: Power supply used for
heating is automatically shut off and an alarm is
given if a sudden graphite tube fracture is detected.
|
|
|